| Animal | Canine |
| Sample type | Serum,plasma |
| Sample volume | 75μL |
| 39-86 °F storage | 24 months shelf life |


Canine NT-proBNP/ Canine Cardiac Troponin I Rapid Quantitative Test is a fluorescence immunoassay used along with Healvet 3000 analyzer for quantitative determination of cNT-proBNP and ccTnI concentration in canine serum or plasma specimen. The test is used as an aid to myocardial function diagnosis and diagnosis of cardiomyocyte injury.
| Animal | Canine |
| Sample type | Serum,plasma |
| Sample volume | 75μL |
| 39-86 °F storage | 24 months shelf life |
| the reference | check details in product description |
Reference range of cNT-proBNP in canine blood
Reference range:
|
Interpretation |
pmol/L |
|
Normal |
<900 |
|
Suspicious, further inspection is recommended |
>900 |
|
Abnormal, further inspection is recommended |
>1800 |
Reference range of ccTnI in canine blood:
Reference range:
|
Interpretation |
ng/mL |
|
Normal |
<0.03 |
|
Suspicion of myocardial injury |
0.03 – 0.12 |
|
High risk of myocardial injury |
>0.12 |
Step 1: Sample Preparation
Add 75 µL of serum or plasma sample into the buffer tube.
Mix thoroughly without forming bubbles.
Step 2: Sample Application
Transfer 75 µL of the prepared mixture into the sample well of the test card.
Step 3: Result Analysis
Insert the test card into the Healvet 3000 Analyzer with standard test mode.
Read and record the test result according to the analyzer instructions.